AUTOCLAVE STERILIZATION

An autoclave is a pressurized device designed to heat aqueous solutions above their boiling point to achieve sterilization. It was invented by Charles Chamberland in 1879. The term autoclave is also used to describe an industrial machine in which elevated temperature and pressure are used in processing materials. It use steam to sterilize microorganism.

Autoclaves are widely used in microbiology, medicine, veterinary science, dentistry, podiatry and metallurgy. The large carbon-fiber composite parts for the Boeing 787, such as wing and fuselage parts, are cured in large autoclaves.


Autoclave have two types :
a. Autoclave with gravitation type.
b. Autoclave with prevaccum type.
The quality of steam is very important for sterilization procces.
A medical autoclave is a device that uses steam to sterilize equipment and other objects. This means that all bacteria, viruses, fungi, and spores are inactivated. However,Prions, like those associated with Creutzfeldt-Jakob disease, also may not be destroyed by autoclaving at the typical 121 degrees Celsius for 15 minutes or 134 degrees Celsius for 3 minutes, but can be destroyed with a longer sterilisation cycle of 134 degrees Celsius for 18 minutes[citation needed].

Autoclaves are found in many medical settings and other places that need to ensure sterility of an object. Many procedures today use single-use items rather than sterilized, reusable items. This first happened with hypodermic needles, but today many surgical instruments (such as forceps, needle holders, and scalpel handles) are commonly single-use items rather than reusable. See waste autoclave.

Because damp heat is used, heat-labile products (such as some plastics) cannot be sterilised this way or they will melt. Some paper or other products that may be damaged by the steam must also be sterilized another way. In all autoclaves, items should always be separated to allow the steam to penetrate the load evenly.

Autoclaving is often used to sterilize medical waste prior to disposal in the standard municipal solid waste stream. This application has grown as an alternative to incineration due to environmental and health concerns raised by combustion byproducts from incinerators, especially from the small units which were commonly operated at individual hospitals. Incineration or a similar thermal oxidation process is still generally mandated for pathological waste and other very toxic and/or infectious medical wastes.

Autoclave quality assurance
Multiple large autoclaves are used for processing substantial quantities of laboratory equipment prior to reuse, and infectious material prior to disposal.
Multiple large autoclaves are used for processing substantial quantities of laboratory equipment prior to reuse, and infectious material prior to disposal.
Sterilization bags often have a “sterilization indicator mark” that typically darkens when sterilization temperatures have been reached. Comparing the mark on an unprocessed bag to a bag that has been properly cycled will show an obvious visual difference.
Sterilization bags often have a “sterilization indicator mark” that typically darkens when sterilization temperatures have been reached. Comparing the mark on an unprocessed bag to a bag that has been properly cycled will show an obvious visual difference.

There are physical, chemical, and biological indicators that can be used to ensure an autoclave reaches the correct temperature for the correct amount of time.

Chemical indicators can be found on medical packaging and autoclave tape, and these change color once the correct conditions have been met. This color change indicates that the object inside the package, or under the tape, has been autoclaved sufficiently. Biological indicators include attest devices. These contain spores of a heat-resistant bacterium, Geobacillus stearothermophilus. If the autoclave does not reach the right temperature, the spores will germinate, and their metabolism will change the color of a pH-sensitive chemical. Physical indicators often consist of an alloy designed to melt only after being subjected to 121°C or 249°F for 15 minutes. If the alloy melts, the change will be visible.

In addition to these indicators, autoclaves have timers, temperature and pressure gauges that can be viewed from the outside.

There are certain plastics that can withstand repeated temperature cycling greater than the 121°C or 249°F required for the autoclaving process. PFA is an example.

Some computer-controlled autoclaves use an F0 (F-naught) value to control the sterilization cycle. F0 values are set as the number of minutes of equivalent sterilization at 121°C or 249°F (e.g: F0 = 15 min.). Since exact temperature control is difficult, the temperature is monitored, and the sterilization time adjusted accordingly.